OPERATION
Carl Zeiss Lighting and contrasting method in transmitted light Axio Lab.A1
74 430037-7144-001 04/2013
(4) Setting the height stop on the condenser
carrier
x Loosen the fastening screw (Fig. 4-2/1) of the
height stop using a SW 3 ball-headed
screwdriver.
x Use focusing drive to focus on the specimen.
x Close luminous-field diaphragm and focus it by
turning the vertical control (Fig. 4-2/2) of the
condenser.
x Carefully raise the condenser slightly without
lifting the specimen.
x Tighten fastening screw (Fig. 4-2/1) of height
stop.
Fig. 4-2 Setting the height stop on the
condenser carrier