Microscope 2
Mag.
Type
Working
distance
(mm)
Numerical
Aperture
(N.A.)
Immersion
medium
Phase ring
position
DIC
position
5x
Plan-Neofluar
13.6
0.15
air
10x
Plan-Neofluar
5.6
0.3
air
20x
EC Plan-Neofluar
2.0
0.5
air
40x
EC Plan-Neofluar
0.21
1.3
oil
63x
Plan-Apochromat
0.19
1.4
oil
100x
Plan-Neofluar
0.2
1.3
oil
Ph3
DIC III
Objectives can be changed using the buttons on the right side of the microscope or through software
by clicking the corresponding icon in MICROSCOPE / COMMON (Figure 9).
Figure 9. Buttons for changing the objective.
5.2 Oil objectives
5.2.1 Changing between oil and air objectives
When you change between objectives, make sure you never change to an air objective while still
having oil on the sample. The air objectives are extremely difficult to clean after being smeared
with oil.