62 Maintenance / Further information
G-30-1439-en OPMI® VISU 200 BrightFlex Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Trouble-shooting table
This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular main-
tenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and con-
tact our service representative.
VISU 200 BrightFlex™ Surgical Microscope
Problem Possible cause See
No function at all. Power cord of suspension
system not plugged in.
Plug in power cord. -
Power switch of suspension
system not on.
Press power switch. Green
light in power switch must
come on.
-
Circuit breaker in suspen-
sion system power switch
activated.
Press power switch again. -
Line power failure. Contact in-house electri-
cian.
-
Lamp module has no con-
tact.
Insert lamp module as far as
it will go.
-
Surgical field illumination at
microscope inoperative.
Illumination system of sus-
pension system not
switched on.
Switch on illumination
system of suspension
system.
-
S light guide not properly in-
serted in lamp or micro-
scope.
Insert S light guide as far as
it will go.
-
Failure of suspension
system electronics.
Illuminate surgical field
using an OR illuminator.
Contact service dept.
-