Axiolab A Carl Zeiss
B 40-015 e 06/99 3-5
3.2.2
Setting of Reflected-Light
Polarization
The Axiolab A microscope is ready for operation as
described in chapter 2, and switched on as
described in section 3.1.
Pull out the plane glass pushrod (3-3/1).
Pull out the dust protection sliders (3-3/2+6)
from the reflected-light illuminator H.
Push polarization slider P (3-3/3) into the slider
mount (vertical slot) from the right.
Push rotary analyzer slider A (3-3/4) in the upper
horizontal slider mount.
For reasons of intensity, no filters
should be contained in the beam path.
Set the analyzer slider to optimum contrast
(maximum dimming) in the field of view via the
adjustment lever.
If possible, this analyzer position should not be changed.
Reduce illumination aperture by turning the adjusting wheel (3-2/4) of the aperture diaphragm
downwards.
To enhance the color contrast of anisotropic object details, the λ-sub compensator slider (3-3/5) can
be inserted in the lower horizontal slider mount.
The color intensity can then be varied via the lever on the λ-sub compensator slider.
3.2.3
Setting of Transmitted-Light Polarization with extended Polarization Equipment
The optional, extended polarization equipment includes the following items:
-
rotary Pol stage (3-4/3)
-
Pol object guide (3-4/4)
-
swing-in polarizer D (3-4/5)
Fig. 3-3 Setting of reflected-light polarization