Operation
Axiostar plus Setting epi-fluorescence Carl Zeiss
B 40-815 e 12/01 3-15
The first epi-fluorescence setting is significantly
easier if you use the Plan-Neofluar 20x/0.50
objective and a strong fluorescent specimen.
Demonstration specimens can be used first.
First leave the illumination beam closed;
move the filter stop carrier 2 (3-10/2) to the
closed position.
Turn on the HBO 50 with the power supply
and let it warm up to operating
temperature for about 15 minutes.
Use the adjustment handle (3-10/3) to
move the desired fluorescence filter
combination on the reflector carrier into the
beam path.
Turn off the halogen lamp (on microscope)
and open illumination beam path. Move the
filter stop carrier 2 to the open position.
Refocus on the specimen and, if necessary,
optimize the collector setting with the knob
(3-10/1) on the HBO 50. Set the collector so
that the illumination from the field of view
appears as equal as possible to the short-
wave excitation from the reflector module.
Correction of the collector position is no
longer necessary for modules with long-
wave excitation.
1 Knob for collector adjustment
2 Filter stop carrier 2
3 Adjustment handle
Fig. 3-10 Setting epi-fluorescence