Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć8
(11) āPositioning/switching on the instrument
• Unless already done, switch on the
instrument using the On/Off switch (2Ć11ā/ā5).
• After switching on, the lamp in the illuminaĆ
tor (2Ć11/4) must light up.
• Set the required brightness via the
illumination control (2Ć11ā/ā6).
• If required, insert attenuation filter or green
filter (2Ć11/3) in the beam path via the filter
slider (2Ć11ā/ā2).
• Move Ph/H/Ph (2Ć11/1), Ph/H/Var (2Ć4/3) or
Var/H/Var sliders to center position.
NOTE If no light appears in the visual
beam path of the Axiovert 25 or
25 CFL, the "observation/camera"
changeover switch (2Ć11/7)
should be activated.
2.2 StartĆup
When starting up the Axiovert 25 microscope for the first time, the instrument must be unpacked
in accordance with section 2.1.1, installed in accordance with 2.1.2, connected to the line and made
ready for operation.
The Axiovert 25 microscope is supplied with factoryĆaligned illumination. This illumination
equipment requires no recentering even after an exchange of lamps by the customer.
The HBO 50 lamp must be centered after it has been changed (see section 3.2 (3)).
The procedure of minor settings, e.g. the adjustment of a ring diaphragm, is explained in the
sections covering the appropriate microscopy techniques.
Otherwise, settings are restricted to changing the objectives (swinging them into the beam path)
and the handling of LD objectives.
Figure 2Ć11 Positioning and switching on the
instrument
1
6
2
3
4
7
5