Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
3Ć8
(4) Change of reflectors Fl
The reflectors are equipped in the factory with the appropriate filter sets consisting of excitation
filter, dichroic beam splitter and barrier filter.
However, it is possible to allow the use of any required Fl filter set.
Two examples of available filter sets:
Blue 450Ć490 (487909) consisting of BP 450Ć490
FT 510
LP 520
Green H 546 (487915) consisting of BP 546ā/ā12
FT 580
LP 590
For further filter sets please see the price list!
Pull reflector mount (3Ć7ā/ā5) out of the Axiovert 25 microscope.
Remove cover (3Ć7ā/ā1).
Loosen two screws (3Ć7/6) using SW 2.5 screwdriver, remove Fl reflector module (3Ć7/2) and
replace it with required, readily equipped reflector module, or insert a new filter set in the
reflector module.
Unscrew screw-on rings (3Ć7/7) in front of the excitation and barrier filter (3Ć7/8).
Unscrew excitation and barrier filter (3Ć7/8) using the suitable wrench, change filters and
secure them using screw-on rings (3Ć7/7); follow the instructions for the installation of
filters!
Loosen screws (3Ć7/9) using SW 2.5 screwdriver and remove the mount for excitation filters
(3Ć7/10).
Change the dichroic beam splitter (3Ć7/11) in spring-mounted frame (3Ć7/12) and attach the
frame to the mount for barrier filters (3Ć7/13); follow the instructions given on the
information sheet included in the packaging.
Attach mount for excitation filters (3Ć7/10) to mount for barrier filters (3Ć7/13) with dichroic
beam splitter and mounting frame and secure it using screws (3Ć7/9).
Attach equipped reflector module to the guide pins (3Ć7/4) in the reflector mount and
secure it using screws (3Ć7/6).
Mark the position of the reflector mount (window is visible when the reflecor mount is pushed
in) using the adhesive label supplied.
Insert the reflector mount in the Axiovert 25 microscope
NOTE The free passage of the reflector mount allows dia. 25 mm filters to be inserted which
are held by a zero ring (3Ć7/3).