OPERATION
Image Acquisition (Acquire) LSM 510
4-56 B 40-51 e 07/98
On the Zoom, Rotation & Offset panel, use the slider to set the desired zoom factor between 1
and 8. This adjustment is infinite.
The scanning frame thus defined is shown by a red outline.
The scanning frame can be rotated and translated at liberty.
Recommended setting Zoom 1
to start with: Rotation 0
Please bear in mind that a reference scale which might be included in the image during the scanning
procedure will only be activated after the end of the scanning procedure if you change the zoom factor
Click on <Channels> in the upper toolbar.
This opens the new dialog boxes Channel
Settings and Excitation of Track, showing
the previously set parameters, which can,
however, be edited.
Optimize the image quality by varying the
pinhole aperture:
A small pinhole diameter will increase the
depth of focus, but reduce the light intensity
received by the PMT.
When you vary the pinhole diameter, an
Optical Slice value is displayed. For optimum
depth resolution, Airy values should be
small, but not below 1.0 to keep the
intensity loss within a reasonable limit.
The parameters
Detector Gain
Ampl. Offset
Ampl. Gain
are described in sections 4.7.1 in the
context of image optimization.
In the Excitation of Track dialog box you can
select other lasers and vary laser intensities.
Fig. 4-56