OPERATION
Image Acquisition (Acquire) LSM 510
4-64 B 40-51 e 07/98
(6) The "Add ROI List" window will appear. Enter any required name to store the ROIs and click on
the <OK> button.
(7) This stored ROI configuration appears in the ROI Lists panel of the "Edit ROI" window.
The following functions are also available in the "Edit ROI" window:
<Close> button
The "Edit ROI" dialog box is closed.
<Remove> button
An entry marked in ROI Lists (stored ROI configuration) is deleted.
<Waste-paper basket> button All the ROIs dragged to the scanning image are deleted. If an area
outline was marked before, this area is now deleted in the
scanning image.
<Auto> button A defined color from the list of colors can be assigned to the ROIs.
In the Auto position, the outlines of the dragged ROIs are
automatically colored differently.
<—> button This button allows you to determine the line thickness of the area
outline.
< > button Clicking on this button in the Edit box allows a ROI to be
deactivated. The tick disappears from the button, as does the
relevant marked area from the scanning image. Clicking on the
button again will reactivate the ROI.
The Use ROI status display in the "Scan Control" window shows whether the ROI mode is activated or
not. When Edit ROI is activated, the ROI mode is active automatically. If ROIs shall not be taken in
consideration during scanning, this button must be used to deactivate the function prior to the scanning
procedure.