OPERATION
Annex LSM 510
4-196 B 40-051 e 07/98
Default settings are 150 and 3000 respectively. With the help of the range indicator the default value is
changed until there is no pixel overflow anymore (‘red pixels’)
Any new value can be set by hand typing and pressing the <ENTER> key while the scan is
running.
Any change in the setting parameters of PMT and diode signal will make a new Gain x in the
ratio formula necessary.
If the adjustment of all parameters is finished only the ratio image can be scanned respectively displayed
by switching off the PMT channel and the diode channel in the "Configuration Control" window and
leaving only the Ratio Channel be turned on. As a result only the ratio image is displayed; which can still
be influenced by the settings in PMT and diode channel.