OPERATION
LSM 510 Image Optimization
B 40-051 e 07/98 4-97
Click on the <Display - Split xy> button of the image processing toolbar.
This displays the separate images scanned in the channels and the composite image.
Fig. 4-105
Now effect image optimization as explained for the single-channel mode, but separately for each
channel.
Now click on the <Display - xy> button.
The composite scan image of two channels is presented in a common window.
Image optimization can be effected much faster if you select a smaller frame, since less data
have to be processed.
Cumulative scan image
Scan image 1 Scan image 2