OPERATION
Image Optimization LSM 510
4-88 B 40-051 e 07/98
On the Channel Settings panel of the "Scan Control" window, set the PMT gain with the Detector
Gain slider.
The image should not have more than a trace of red.
Adjust the Ampl. Offset slider so that areas without picture content just show a trace of blue.
If necessary, re-amplify brightness with the Ampl. Gain slider.
☞
Do not change the Ampl. Gain setting unless the settings made so far are insufficient for
optimizing the image.
Fig. 4-97
In the Color Palette List panel of the "Color Palette" window click on "No Palette".
This deselects the "Range Indicator" and activates the new presentation.
In the "Scan Control" window click on the <Stop> button.
This stops the continuous scan.
Commentaires sur ces manuels