OPERATION
LSM 510 Image Acquisition (Acquire)
B 40-051 e 07/98 4-29
To switch on the Enterprise laser, proceed as follows:
(1) The internal water cooling LP 5 is running.
(2) Start the PC, wait until NT system is booted.
(3) Switch on the Power Supply of the Enterprise laser, power potentiometer turned to maximum.
(4) Start the LSM 510 software.
If the LSM 510 software is already running and you want to use the UV laser, do the following:
(1) Close the LSM 510 software.
(2) Switch on the power supply of the Enterprise / power potentiometer turned to maximum.
(3) Start the LSM 510 software again.
This applies to HeNe lasers:
After selecting the laser, click on the <On> button.
After a short heat-up phase of about 15 seconds, "Status ready - On" appears.
The required laser for image acquisition is now available.
Click on the <Close> button to close the "Laser Control" window.