OPERATION
LSM 510 Annex
B 40-051 e 07/98 4-185
4.12.6 Specification of Trigger-Interface LSM510
Application:
With the LSM510 Release 2.01 you can control various actions externally using Trigger-In or force
external devices to work at a defined time depending on an action using Trigger-Out during time series.
These actions are: Scan-Start/Stop, Bleach, Change of Scan-Interval, end of a countdown or even a
mouse-click on a button.
Interface:
Front plate Scanner-Interface (Scan-IF) inside
Elektronic-Box (Scan-Control-Module) of LSM510:
Connector ‘User I/O’, 26-pin shrinked SUB-D
Number:
4x Trigger-In, 4x Trigger-Out
Type/Voltage Range:
TTL (HCMOS), 0.0 - 5.0 V
Load:
In: 22 kOhm input impedance
Out:± 4 mA