OPERATION
Image Acquisition (Acquire) LSM 510
4-36 B 40-051 e 07/98
Other filter sets:
The filter sets described in this section are included in the standard configuration, but other sets are
available on request.
DAPI: BP 365 FSET01
FT 395
LP 397
TRITC: BP 546 FSET15
FT 580
LP 590
Use the focusing drive (4-37/5) to focus the required object plane.
Select specimen detail by moving the stage in X and Y via the XY coaxial drive (4-37/6 and 7).
This completes the conventional setting of the microscope before starting the laser scanning of the
specimen.
Click on the <Close> button to close the "Microscope Control" window.